Calibration procedure of Differential pressure (DP) level transmitter

Requirement:-
Data sheet
Pressure calibrator
Multimeter
Hart communicator
Procedure:-
1) Put the controller in manual mode for control loop.
2) Connect HART Communicator and verify some parameters by refer to data sheet. Typical parameters are, tag number, PV, LRV and URV.
3) Isolate the instrument from the process.
4) Release both pressure and drain low and high side liquid throughout manifold drain.
5) Open both plug at seal pot
6) Top up seal liquid for low side until half of seal pot.
7) Remove tubing at high side of manifold
8) Hook up pressure calibrator at high side and isolate the vent valve
9) Hook up a multimeter in series with the signal to the DCS to measure current signal.
10) Apply pressure as per calculation when level equivalent to zero
11) Multimeter should show 4mA
12) If not, do zero adjustment at transmitter using HART Communicator
13) Apply pressure as per calculation when level equivalent to 100%Multimeter should show 20mA
If not, do span adjustment at transmitter using HART Communicator
14) After completion of the job ask panel operator to put loops back in normal mode.
15) Fill the calibration form and file it for future reference.
Note:There is many way to perform calibration for DP level but the important thing is the DP reading at HART communicator should same with the LRV and URV
Example:-
Product S.G=0.89
Glycol S.G = 1.02
A= 2000mm (measurement length)
B= 100mm (off set)
C=A+B=2100mm
Dp = pressure at high side – pressure at low side
LRV= (B x Product S.G) – (A x Glycol S.G)
= (100mm x 0.89) – (2000 x 1.02)
= 89 – 2040
= -1951 mmH2O
URV = (C x Product S.G) – (A x Glycol S.G)
= (2100mm x 0.89) – (2000 x 1.02)
= 1869 – 2040
= -171 mmH2O
Note:
Calculation for apply pressure at high side manifold transmitter
0% = (B x Product S.G)
= (100mm x 0.89)
= 89 mmH2O
100%= (C x Product S.G)
= (2100mm x 0.89)
= 1869 mmH2O

Calibration Procedure for Differential Pressure (DP) Flow Transmitter

Requirement:-
Data sheet
Pressure calibrator
Multimeter
Hart communicator
 
Procedure:-
1) Put the controller in manual mode for control loop.
2) Connect HART Communicator and check parameters by refering to data sheet. Typical parameters are, tag number, PV, LRV and URV.
3) Isolate the instrument from the process.
NOTE – If the process is hazardous, please unsure proper flushing is done to remove the entire hazard.
4) Release both pressure and drain low and high side liquid throughout manifold drain.
5) Connect pressure calibrator to high side of manifold
6) Expose the low side to atmosphere
7) Connect multimeter in series with the signal to the DCS to measure current signal.
8) Apply pressure as per data sheet LRV (normally 0 mmH2O)
9) Multimeter should show 4mA
10) If not, do zero adjustment at transmitter using HART Communicator
11) Apply pressure as per data sheet URV
12) Multimeter should show 20mA
13) If not, do span adjustment at transmitter using HART Communicator
14) Verify the linearity by increasing and decreasing the pressure (0%,25%,50%,75%,100%,75%,50%,25% and 0%of range)
15) After completion of the job ask panel operator to put loops back in normal mode or normalize the MOS
 
Example:-
Instrument calibrated range: 0 – 2500 mmH2O
Process calibrated range: 0 – 105000 Nm3/h
Q =K  P
0%  0 mmH2O= ( 0/2500) x 105000
= 0 Nm3/h 
50%  1250 mmH2O= ( 1250/2500) x 105000
= 7424.6 Nm3/h 
100%  2500 mmH2O= ( 2500/2500) x 105000
= 105000 Nm3/h

Calibration of Control Valve

Requirement:-
Data sheet
HART calibrator
 
Procedure
1) Put the controller in manual mode for control loop 
 
2) Isolate the Control valve from the process.
NOTE – Careful step shall be done to ensure no upset to the operation.
 
3) Connect HART Communicator and check parameters by refering to data sheet such as Tag Number, OP and etc. 
 
4) Change from normal mode to setup mode
 
5) Select auto find stop (wait some minute)
 
6) If full open is more than 100% select open adjustment and put new number as per calculation
 
7) Make an auto calibration ( wait some minute)
 
8) Verify the linearity by increasing and decreasing the travel (0%,25%,50%,75%,100%,75%,50%,25% and 0%of range)
 
9) After completion of the job put loops back in normal mode

Calibration procedure DP transmitter

How to calibrate Differential pressure transmitter (capillary type)?

Requirement:-
Data sheet
Pressure calibrator
Multimeter (for mA)
Hart communicator

Procedure:-
1) Put the controller in manual mode for control loop.

2) Connect HART Communicator and check parameters like tag number, PV, LRV and URV.

3) Isolate the instrument from the process.
NOTE – If the process is hazardous, please unsure proper flushing is done to remove the entire hazard.

4) Open vent valve at drip ring and open plug at the top of the drip ring (if available) to release the process pressure

5) Clean the liquid inside the drip ring and put back the plug for high side only (if available).

6) Expose the low side to atmosphere.

7) Now, Connect a multimeter in series with the signal to the DCS to measure current signal.

8) In this condition both capillary flange will be atm pressure at difference high

9) Multimeter should show 4mA and PV at HART communicator should shows 0 (LRV)

10) If not, do zero adjustment at transmitter using HART Communicator

11) Connect pressure calibrator to high side flange (drip ring)

12) Apply pressure depend data sheet span (URV)

13) Multimeter should show 20mA and PV at HART communicator should shows same with URV

14) If not, do span adjustment at transmitter using HART Communicator

15) Verify the linearity by increasing and decreasing the pressure (0%,25%,50%,75%,100%,75%,50%,25% and 0%of range)

16) After completion of the job again put loops back in normal mode.

Calibration

Overview

We offer expert Instrument Calibration & Repair Services for all types of calibration and test & measurement instruments. Extensive metrology experience is applied to every calibration we perform, and our strict adherence to our quality registrations assures you of complete compliance to the standards that affect your industry.

Calibration services are designed to help the manufacturers and users of instruments achieve the highest possible levels of measurement quality and productivity.

Our laboratory is equipped to service thousands of electronics test & measurement instruments. Our quality system guarantees you the best service available, fast turnaround, and competitive pricing.

Calibration! What is it ?

A set of operations, performed in accordance with a definite procedure, that compares the measurements performed by an instrument to those made by a more accurate instrument or a standard for the purpose of detecting and reporting, or by adjustment, errors met in the instrument tested.

Purpose of Calibration

To ensure readings from an instrument are consistent with other instruments and to determine the accuracy of the instrument i.e. that it can be trusted for its observed/displayed measured value.

The value of calibration for industry

To ensure that products are manufactured to specifications.
To demonstrate that the industry operates a quality system and technically competent and are able to generate technically valid results.
To increase quality & value of product.
The calibrated measuring instruments (working standards) have the assurance of an unbroken chain of national/international measuring standards.

What is Traceability ?

Property of the result of a measurement or related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties.

Why measurements must be Traceable ?
Traceable measurements ensure the uniformity of manufactured goods and industrial processes.
To support equity in trade as well as compliance to regulatory laws & standards.
Essential to the development of technology.
To provide traceable calibration of test and measurement equipment is an ISO-9000 requirement(Purpose: With an eye to provide safety and fitness for use) to ensure that the products manufactured in one country will be acceptable in another on the basis of agreed to measurement standards, methods & practices)
Provides users with assurance of the confidentiality, validity and accuracy of the data provided while testing.

Uncertainty- How accurate are your measurements ?

It is a measure of the quality of a measurement & provides the means to assess & minimize the risks and possible consequences of poor decision. (A measurement with a stated uncertainty can be compared to others)

Measurements are traceable when they can be related to recognised measurement system through an unbroken chain of comparisons each with stated uncertainties.

Calibration Interval-Validity Period.
Validation is usually nominal maximum period between two successive calibrations. The validity period is generally considered to be maximum appropriate, provided.
The equipment is of good quality and is of proven adequate stability.
If any suspicion or indication of overloading or mishandling arises, the equipment will be checked immediately and thereafter at frequent intervals until it can be shown that stability has not been impaired.
The extension of validity period is based on factors such as history of stability, frequency of use, accuracy required, ability of staff to perform regular checks and successful participation in proficiency testing programs. It is the responsibility of the testing authority to provide evidence that its calibration system ensures that confidence in the equipment can be maintained.
Adequate records of the test data must be maintained over a period of time to draw support for extension of validity period.

Calibration Measurement! How ?

Verifying the performance of the test Instruments which involves applying a standard value to the Instruments and observing how it behaves.

Adjust the response of the Instruments, wherever possible(After verifying response, reset, adjust/correct the Instruments via internal control which in normal course is not accessible)

Obtain correction factor for the test Instruments.(Accordingly correction factors/calibration constants are calculated & incorporated in measurements)

Posted from WordPress for Windows Phone